Volumetric infusion pump AP31P
Simple, safe & reliable
Designed to meet the needs of modern I.V. therapy in ICU, CCU,
anesthesia, surgery, cardiology and oncology.
Key features:
extremely simple handling
wide range of infusion rate
easy modification of infusion parameters in
running course
comprehensive system of audible and visual
alarms & warnings
free-flow blocking device
long battery life
3 modes of operation: flow rate, flow rate
and volume, flow rate and time
|
Performances |
|
|
Flow rate range: |
1 to 1000 ml/h
(programmable in 1 ml/h steps) |
|
Volume to be infused:
|
1 to 10 000 ml
(programmable in 1 ml steps) |
|
Infusion time:
|
up to 99 h 59 min 59
sec |
|
Setting modes: |
flow rate only, flow
rate and volume, flow rate and time |
|
KVO rate: |
1 ml/h for flow rate
1 - 10 ml/h
2 ml/h for flow rate 11 - 100 ml/h
3 ml/h for flow rate 101 - 300 ml/h
4 ml/h for flow rate 301 - 500 ml/h
5 ml/h for flow rate 501 - 1 000 ml/h |
|
Flow rate accuracy: |
± 5% |
|
Occlusion pressure
limit: |
80 ± 20 kPa (0,8 ±
0,2 kG/cm2) |
|
Compatible I.V. set: |
only special
(dedicated) ASCOSETâ Peristaltic infusion sets with silicone insertion
can be applied |
|
Extended system of
alarms and warnings (both audible and visual): |
Alarms: occlusion,
end of infusion, KVO infusion, air in line, open door, no mains,
discharged batteryPrealarms: 5 min left to the end of infusion, low
battery
Protection against free flow: special free-flow blocking device
protects the patient against any uncontrolled infusion in case of the
pump door opening (e.g. during mounting or replacement of I.V. set) |
|
|
>>back
|
|
Technical
specifications: |
|
|
Pumping system: |
peristaltic finger |
|
Power supply:
|
230 V AC ± 10%, 50/60
Hz, max. 5 VA or from a built-in battery |
|
Built-in battery: |
rechargeable, NiCd
battery of increased durability, automatically charged when the pump is
connected to AC power supply |
|
Battery life: |
min 4 h at a flow rate
1 000 ml/h
24 h at a flow rate 5 ml/h |
|
Battery charging
time: |
24 h |
|
Electrical safety: |
Class II, type BF (EN
60601-1), IP31 (splash-proof) |
|
Air bubble detector: |
ultrasonic, air in
line alarm with air bubbles of 0,05 ml in size |
|
Drop detector: |
optical, with digital
disturbance filtration |
|
Weight: |
3,6 kg |
|
Dimensions (WxDxH): |
140 x 230 x 230 mm |
|
Operating
temperature: |
5 - 40° C |
|
Pump fixation: |
AP31P can be placed
onto a table, or secured to a pole or bed-frame with the incorporated
clamp |
|
|
>>back
|
|
|
|