Single-syringe pump AP12
Simple, safe & reliable
Designed to meet the needs of modern I.V. therapy in
neonatal/pediatric/adult ICU, CCU, anesthesia, cardiology and oncology.
Key features:
user friendly software simplifies set-up
and operation
wide range of infusion rate
over 40 types of compatible syringes (10-60
ml)
automatic & manual bolus capability
automatic reduction of residual bolus after
occlusion release
easy modification of infusion parameters in
running course
comprehensive system of audible and visual
alarms & warnings
drug library with user selectable drug
names
long battery life
data lock with a software-controlled
passkey system
4 modes of operation: flow rate, flow rate
and volume, flow rate and time, volume over time
|
Performances |
|
 |
Flow rate range: |
0.1 - 500 ml/h for 50
ml syringe
0.1 - 300 ml/h for 30 ml syringe
0.1 - 250 ml/h for 20 ml syringe
0.1 - 150 ml/h for 10 ml syringe
(0.1 - 99.9 ml/h in 0.1 ml/h steps; 100 - 500 ml/h in 1 ml/h steps) |
|
Volume to be infused:
|
0.1 - 1000 ml
(0.1 - 99.9 ml in 0.1 ml steps; 100 - 1000 ml in 1 ml steps) |
|
Maximum bolus rate:
|
2000 ml/h for 50 ml
syringe
1200 ml/h for 30 ml syringe
1000 ml/h for 20 ml syringe
600 ml/h for 10 ml syringe |
|
Bolus dose /
Induction doeses:
|
limited to syringe
size |
|
Infusion time
selection: |
up to 100 h |
|
KVO rate: |
0.1 - 5 ml/h,
programmable in 0.1 ml/h steps |
|
Flow rate accuracy: |
± 2% (complies with
EN 60601-2-24) |
|
Occlusion pressure
limit:
|
4 - level selection:
high |
120 kPa ± 30 |
medium |
90 kPa ± 20 |
low |
60 kPa ± 30 |
minimum |
40 kPa ± 30 |
|
|
Compatible syringes: |
10, 20, 30 and 50/60
ml |
|
Syringe brands: |
B.BRAUN, BD,
MONOJECT, CODAN, TERUMO, IVAC, INJECTOMAT, PENTAFERTE and other (see
operating manual) |
|
Extended system of
alarms and warnings
(both audible and visual): |
Alarms: no syringe,
wrong syringe, empty syringe, end of infusion, occlusion, no mains,
discharged battery, internal malfunctionPrealarms: 5 minutes left to
the end of infusion or till the syringe gets empty, low battery |
|
|
>>back
|
|
Technical
specifications: |
|
|
Power supply:
|
230 VAC ± 10%, 50/60
Hz, max. 10 VA or from a built-in rechargeable battery |
|
Power consumption:
|
max. 10 VA |
|
Built-in battery: |
rechargeable, NiCd of
increased durability, automatically charged when the pump is connected
to AC power supply |
|
Battery life: |
24 h at a flow rate of
5 ml/h |
|
Battery charging
time: |
24 h |
|
Electrical safety: |
Class I, type BF,
IP31 |
|
Conformity standards:
|
EN 60601-1
EN 60601-1-2 (EMC)
EN 60601-2-24
MDD 93/42/EEC |
|
Weight: |
3,1 kg |
|
Dimensions (WxDxH): |
320 x 180 x 140 mm |
|
Operating conditions:
|
5 - 40° C, relative
humidity: 20 - 90% |
|
|
>>back
|
|
Other
features: |
|
|
External alarm
socket: |
24 V, 1 A |
|
Other functions: |
Anti-bolus system (ABS)
Automatic syringe size recognition
Battery life indicator
User selectable drug names
Programmable name of ward
Infusion pressure monitoring
Infusion parameters protected by a
password
Priming
STAND-BY
Extended menu of user tests and
service tests
|
|
|
>>back
|
|
|
|